Ecotest MKS-05 Terra/P – Dozimetre

Specifications

Measurement ranges and main relative permissible error limits
Gamma and X-ray radiation ambient dose equivalent rate (137Cs) µSv/h 0.1…9 999;
% ±(15+2/H*(10)), where  H*(10) is a numeric value of measured DER equivalent to µSv/h
Gamma and X-ray radiation ambient dose equivalent (137Cs) mSv 0.001…9 999;
±15%
Beta-particles flux density (90Sr+90Y) 1/(cm²×min) 10…100 000;
% ±(20+200/φβ), where φβ is a numeric value of measured beta-particles flux density equivalent to part./(cm²×min)
Ambient dose equivalent accumulation time 1 min…9 999 h
Energy ranges of measurement and energy dependence:
Gamma and X-ray radiation MeV 0.05…3.0; ±25%
Beta radiation MeV 0.5…3.0
Resolution of threshold level programming for:

  • – dose rate
  • – dose
  • – flux density
µSv/h
mSv
10³/(cm²×min)
0.01
0.001
0.01
Battery life* hours 2 000
Operating temperature range °С -20…+50
Weight kg 0.2
Dimensions mm 55×26×120